Peptide purity: HPLC and mass spectrometry explained
"98% pure" is a useful number only once you know what was measured, how, and what the method can't see.
Purity and identity are different questions
Identity asks whether the main component is the intended sequence. Purity asks how much of the detectable material is that main component. A sample can be highly pure and still the wrong peptide. So a sound COA answers both questions, usually with HPLC for purity and mass spectrometry for identity.
How reversed-phase HPLC purity works
In reversed-phase HPLC the sample passes through a hydrophobic column, typically C18. Increasing organic solvent (for example acetonitrile with an acid modifier such as TFA) elutes components at different times. A UV detector, commonly at 214–220 nm, records each peak. Purity is the main peak's area as a percentage of the total integrated area.
The result has limits. Impurities that co-elute with the main peak, or that don't absorb at the detection wavelength, are under-counted. Results can also vary with the column, the gradient and how peaks are integrated. That is why the method details matter.
What mass spectrometry adds
MS measures mass-to-charge ratios. Comparing the observed molecular mass with the theoretical mass supports identity, and can reveal common synthesis-related impurities such as deletion sequences, incomplete deprotection or oxidation, which change the mass. LC-MS, which couples the two techniques, can assign masses to individual HPLC peaks.
Net peptide content
Lyophilised peptides contain bound water and counter-ions as well as the peptide. Net peptide content, determined for example by amino acid analysis or elemental nitrogen, gives the fraction of powder mass that is peptide. It is often well below the HPLC purity figure, and it matters when you prepare solutions of a known concentration for an assay.
Related
- HPLC vs LC-MS for peptides: identity and purity workflow
- Peptide purity vs peptide content: a vial mass balance
Further reading
- Reversed-phase HPLC of peptides: the basics, with a worked area-percent example
- How peptides are made: solid-phase synthesis
Sources and further reading
- United States Pharmacopeia, General Chapter <621> Chromatography.
- ICH Q2(R2), Validation of Analytical Procedures.
- ICH Q6B, Specifications for Biotechnological/Biological Products.
Frequently asked questions
What does a purity percentage measure?
The main peak's area as a percentage of the total integrated area in a reversed-phase HPLC run with UV detection, usually at 214 to 220 nm. It depends on the column, gradient and integration.
Can a peptide be highly pure and still be the wrong peptide?
Yes. Purity asks how much of the detectable material is the main component; identity asks whether that component is the intended sequence. A sound COA answers both.
What does mass spectrometry add?
It measures mass-to-charge ratios, so the observed mass can be compared with the theoretical mass and common impurities such as deletions, incomplete deprotection or oxidation can show up as mass shifts.
Why is net peptide content lower than purity?
Lyophilised peptides also contain bound water and counter-ions, so the fraction of powder mass that is peptide is often well below the HPLC purity figure.