Analytical methods
Reversed-phase HPLC of peptides: the basics
Nearly every peptide purity figure on a Certificate of Analysis comes from one technique. This is how it works and what the number can and cannot tell you.
What the instrument does
High-performance liquid chromatography has been used extensively for the isolation and purification of peptides; Mant et al. (2007) cover the major modes used for peptides (size-exclusion, ion-exchange and reversed-phase). For purity testing of synthetic peptides, reversed-phase is the workhorse. A small volume of dissolved peptide is injected into a solvent stream that passes through a column packed with hydrophobic particles, commonly C18-modified silica. Peptides stick to the column in a mostly aqueous mobile phase and are released as the proportion of organic solvent, typically acetonitrile, is raised in a gradient. A UV detector at the column outlet records absorbance against time.
The pieces of a method
| Part | Typical choice | What it controls |
|---|---|---|
| Column | C18 (or C8, C4 for larger or more hydrophobic peptides), narrow bore, small particles | How strongly peptides are retained and how well neighbouring species resolve. |
| Mobile phase A | Water with an acid modifier, such as 0.1% TFA | Starting retention. Krokhin and Spicer (2009) describe C18 sorbents with trifluoroacetic or formic acid as the ion-pairing modifier as two of the most popular modes. |
| Mobile phase B | Acetonitrile with the same modifier | Elution strength as its proportion rises. |
| Gradient | A programmed rise in %B over many minutes | Separation: a shallower gradient generally spreads closely related peptides further apart. |
| Detection | UV near 214 to 220 nm (peptide bond); 280 nm for Trp and Tyr | Which species are counted. A compound that does not absorb at the chosen wavelength is invisible. |
Krokhin and Spicer (2009) propose a hydrophobicity index for peptides: the acetonitrile concentration that gives a retention factor of 10 under isocratic elution. They describe it as a characteristic of the peptide-sorbent interaction that depends only on the sorbent type and the ion-pairing modifier. The practical point is that a retention time is only meaningful together with the column, gradient and modifier that produced it.
From chromatogram to a purity percentage
The data system integrates each peak and reports area. Purity is the main peak area divided by the total integrated area. The worked example below uses an invented chromatogram, not data from any product of ours.
| Peak | Retention time (min) | Area (arbitrary units) | Area % | Typical interpretation |
|---|---|---|---|---|
| 1 | 3.1 | 0.6 | 0.6% | Early peak, usually small polar species or injection-related |
| 2 | 9.2 | 1.3 | 1.3% | Elutes before the main peak (for example a deletion sequence) |
| 3 | 9.9 | 95.4 | 95.4% | Main peak |
| 4 | 10.5 | 1.5 | 1.5% | Elutes just after the main peak (for example an oxidised or protecting-group-bearing form) |
| 5 | 12.8 | 1.2 | 1.2% | Late, more hydrophobic species |
| Total | 100.0 | 100.0% | Purity = main peak ÷ total = 95.4% |
The example shows two points. First, the figure is a ratio of peak areas, so the integration thresholds matter: a small peak the software does not integrate does not lower the number. Second, the 4.6% of area outside the main peak is not necessarily peptide; it can be anything that absorbs at that wavelength.
Where the number can mislead
- Co-elution. Two species can share a peak. Mass spectrometry across the peak or a second method with different selectivity can reveal this (checking the mass spectrometry).
- Different absorbance. Area percent assumes all species respond equally at the wavelength used. Species lacking the chromophore or with different composition respond differently.
- Not a content measurement. Water, TFA and other counter-ions are mostly invisible at 214 nm and do not appear in the purity figure at all (purity vs content; amino-acid analysis).
- Method dependence. The same sample can give different purity numbers on different gradients or columns, which is why a COA should state the method.
What a useful COA shows
- The chromatogram itself, not just the percentage.
- Column, mobile phases, gradient and detection wavelength.
- Retention time of the main peak and the date and batch it refers to.
- A mass-spectrometry result for the same batch to cover identity.
See how to read a peptide COA and peptide purity: HPLC and mass spectrometry explained. Our product pages label stated purity as supplier-stated and show the batch COA as not yet provided until one exists.
Related
- How to read a COA HPLC chromatogram
- Solid-phase peptide synthesis explained
- Common impurities in synthetic peptides
- TFA and other counter-ions
References
- Mant CT, Chen Y, Yan Z, et al. (2007). HPLC analysis and purification of peptides. Methods Mol Biol 386:3-55. PubMed 18604941 · DOI
- Krokhin OV, Spicer V (2009). Peptide retention standards and hydrophobicity indexes in reversed-phase high-performance liquid chromatography of peptides. Anal Chem 81:9522-30. PubMed 19848410 · DOI
Frequently asked questions
How does reversed-phase HPLC separate peptides?
Peptides bind to a hydrophobic stationary phase (commonly C18) and are eluted by an increasing concentration of organic solvent such as acetonitrile, with an acid modifier such as TFA or formic acid; more hydrophobic peptides elute later (Mant et al., 2007; Krokhin and Spicer, 2009).
Why is TFA used in the mobile phase?
It acts as an ion-pairing modifier that sharpens peptide peaks. Krokhin and Spicer (2009) note that retention depends on the sorbent and on the ion-pairing modifier used, which is why TFA and formic acid methods give different retention.
How is HPLC purity calculated?
As the main peak's area divided by the total integrated area, as a percentage. The worked example below shows the arithmetic.
What wavelength is used?
Peptide analyses commonly monitor around 214 to 220 nm, where the peptide bond absorbs.
Can HPLC purity be misleading?
Yes. A species that does not absorb at the chosen wavelength or that co-elutes with the main peak is not counted separately, and area percent says nothing about water or counter-ion content.