Documentation
How to read a peptide Certificate of Analysis
A Certificate of Analysis (COA) records the tests a specific batch passed. Here is what each part means and what to check before you rely on one.
1. Header: who issued it and for which batch
A COA should name the issuing organisation, the product and a batch or lot number, along with the test or release date. The lot number is what ties the document to the material in your vial, so check that it matches the label. A COA with no lot number, or a lot number that doesn't match, tells you nothing about your material.
2. Identity: is it the right molecule?
Identity is usually confirmed by mass spectrometry (MS), such as ESI-MS or MALDI-TOF. The COA compares the observed mass with the theoretical mass calculated from the sequence. A close match supports identity. Note which ion is reported (for example [M+H]+ or multiply charged ions), because that changes the expected value.
3. Purity: how much of the sample is the target peptide?
Purity is typically measured by reversed-phase HPLC with UV detection near 214–220 nm, where the peptide bond absorbs. It is reported as the target peak's area as a percentage of all detected peak area. A good COA includes the chromatogram, or at least the column, gradient and wavelength. See our HPLC and MS explainer for what this figure does and doesn't capture.
4. Appearance, solubility and other tests
COAs often record appearance (for example "white lyophilised powder"). They may also record water content (Karl Fischer), counter-ion content (for example acetate or TFA) or net peptide content. Net peptide content is not the same as HPLC purity: a lyophilised powder also contains water and counter-ions, so the peptide mass per vial is lower than the gross powder mass.
5. Red flags
- No lot number, or one that doesn't match the vial.
- A purity figure with no stated method or chromatogram.
- Identity "confirmed" with no observed mass value.
- The same document reused across different batches or products.
- No named issuing laboratory, signature or date.
Further reading
- How to read a COA HPLC chromatogram
- TFA to acetate: peptide counter-ion exchange
- N- and C-terminal modifications and mass spectrometry
- HPLC vs LC-MS for peptides: identity and purity workflow
- Reversed-phase HPLC of peptides: the basics
- Lyophilized peptides: what freeze-drying does and why
Sources and further reading
- United States Pharmacopeia, General Chapter <621> Chromatography.
- ICH Q2(R2), Validation of Analytical Procedures (International Council for Harmonisation).
- ICH Q6B, Specifications: Test Procedures and Acceptance Criteria for Biotechnological/Biological Products.
Frequently asked questions
What should a peptide COA contain?
The issuing organisation, the product and a lot number that matches your vial, an identity result with the observed mass, a purity result with its method, and often appearance, water, counter-ion or net peptide content.
How do I know a COA belongs to my vial?
Check that the lot number on the document is the lot number on the label. A COA with no lot number, or a different one, tells you nothing about your material.
Which test confirms identity?
Usually mass spectrometry, such as ESI-MS or MALDI-TOF, comparing the observed mass with the mass calculated from the sequence. Note which ion is reported, because that changes the expected value.
What are the red flags on a COA?
No lot number or a mismatched one, a purity figure with no method or chromatogram, identity stated without an observed mass, one document reused across batches or products, and no named laboratory, signature or date.