For laboratory research use only. Not for human or veterinary use. Read our research-use policy

Laboratory practice

Peptide adsorption to vials and plastics

Dilute peptide solutions lose material to the tubes, tips and vials that hold them. This page collects what the published studies measured, shows the arithmetic of why concentration matters, and screens 18 peptides from our library by sequence.

Published 19 November 2026 · Last updated 19 November 2026

The short version

Peptides adsorb to the surfaces they touch. Lindberg and Andersson (1991) found measurable adsorption of atrial natriuretic peptide to all of the materials they tested, including siliconized glass and polypropylene, and showed that adding albumin or Triton X-100 prevented or partly reversed it. Maes et al. (2014) describe nonspecific binding at pipette tips, sample tubes and LC vials, and at autosampler, column and mass spectrometer components, and note that because peptide properties vary so widely there is no single general fix. The consequence for a laboratory is practical: a concentration you calculated from the vial label is not the concentration in the tube if the solution is dilute and the surface is unfavourable. See our guides to net peptide content and reconstitution volume for the weighed side of the same question.

Why concentration changes the picture

Suppose a surface removes a fixed amount of peptide, whatever the concentration. The fraction lost then depends on how much peptide is in the solution to begin with. The table uses 1 pmol as an illustrative fixed loss from 1 mL; the number is arithmetic for illustration, not a measured property of any surface or peptide.

Fraction lost if a surface takes a fixed 1 pmol from 1 mL (illustrative arithmetic)
ConcentrationPeptide in 1 mLFraction lost
1 nM1 pmol100%
10 nM10 pmol10%
100 nM100 pmol1.0%
1 µM1,000 pmol0.100%
100 µM100,000 pmol0.001%

The measured case is similar in kind. In Lindberg and Andersson's experiment, 1 mL of a 120 pM solution was dispensed in turn into six empty polystyrene tubes and 73% of the starting peptide was lost to adsorption; with 1 mL at concentrations from 1 nM to 10 µM, 10 to 31% adsorbed onto a 15 cm² polystyrene surface, roughly the area of a conventional test tube.

What the studies measured

Published adsorption studies and what each reports
StudySystemWhat the authors report
Lindberg and Andersson (1991)Atrial natriuretic peptide in polystyrene, polypropylene and siliconized glass, by radioimmunoassayAdsorption to every material tested; prevented or partly reversed by adding human serum albumin or Triton X-100.
Siebels et al. (2024)Reference mixture of thousands of tryptic peptides in vials of different surface chemistry, by LC-MS/MSA significant number of hydrophobic peptides adsorbed to conventional polypropylene; few adsorbed to low-protein-binding polypropylene; the most adsorbed to glass, driven by electrostatic interactions.
Kune et al. (2025)HeLa tryptic digest standards in different vial materials and with additivesPreferential adsorption of hydrophobic peptides onto polypropylene vials; vials made from polymers with polar monomer units (PMMA or PET) reduced the loss, and nonionic detergents at 0.0001% (PEO) and 0.0075% (DDM) improved performance.
Zhang et al. (2022)50 BSA digest peptides in LoBind and ordinary polypropylene tubes, polypropylene and glass vials, and across LC conditionsHydrophobicity index and peptide length correlated with adsorption but explained about 30% of the peptides; modified polypropylene gave recovery above 80% within 24 hours; a C8 column, strong needle wash and slow gradient reduced carryover.
Zhou et al. (2015)Ten neuropeptides in microdialysis with capillary LC-MSAcetonitrile in the sample eliminated carryover; treating a negatively charged membrane with polyethylenimine improved recovery only for peptides with net positive charge.
Bongaerts et al. (2023)Neuromedin U-8 (8 residues) and neuromedin S (36 residues)Larger neuropeptides (23 to 36 residues) were harder than those under 15 residues; adding 0.05% plasma as an adsorption competitor was found to be essential.
Vanderstichele et al. (2016)Amyloid-β peptides in CSF in different tubes, by ELISALow-binding tubes reduced loss on aliquoting and across freeze-thaw cycles, and the Aβ1-42/Aβ1-40 ratio was less affected by tube type than Aβ1-42 alone.
Hamada et al. (2026)Liraglutide with silicone oil, headspace and agitation, by SEC, CD, ThT, Raman and NMRPeptide adsorption on a silicone oil surface was imaged; the authors conclude that combined air-water and silicone-oil-water interfacial stress, not either alone, drives liraglutide aggregation.

Three themes recur: surface chemistry (hydrophobic polypropylene and polystyrene, charged glass), peptide properties (hydrophobicity, charge and length), and the solution (organic solvent, competitors, detergent).

A screen for 18 library peptides

The table below applies the literature's recurring properties to the sequences on our charge and pI page. The thresholds are our own choices for a first look: a hydrophobic share of 40% or more, a computed net charge of +2 or more at pH 7.4 (a candidate for electrostatic binding to negatively charged surfaces such as glass), and 23 or more residues. Zhang et al. (2022) found that sequence properties explain only part of the variation, so a peptide with no flag is not guaranteed to be safe, and a flagged one may behave well. 7 of the 18 carry at least one flag.

Where adsorption literature would suggest extra care, for 18 peptides in our library (a screen computed from sequence, not a measurement)
PeptideResiduesHydrophobic share (A, I, L, M, F, V, W)Net charge at pH 7.4Flags
BPC-1571527%−2.2none flagged
KPV333%+0.8none flagged
Selank70%+1.8none flagged
Semax729%−1.1none flagged
DSIP933%−2.2none flagged
Epithalon425%−2.2none flagged
PE 22-28743%+0.8hydrophobic share
Pinealon30%−1.2none flagged
Vesugen30%−1.2none flagged
Kisspeptin-101040%+1.8hydrophobic share
VIP2836%+3.9cationic at pH 7.4; long chain
Oxytocin922%+0.8none flagged
ARA-2901127%−2.0none flagged
AOD-96041631%+0.8none flagged
MOTS-c1638%+2.8cationic at pH 7.4
Humanin2446%+1.7hydrophobic share; long chain
Thymosin alpha-12829%−6.0long chain
LL-373735%+5.8cationic at pH 7.4; long chain

What the literature supports doing

  1. Match the container to the peptide. For dilute solutions of hydrophobic peptides, low-protein-binding polypropylene or vials made from polar polymers performed better than conventional polypropylene (Siebels et al., 2024; Kune et al., 2025; Zhang et al., 2022).
  2. Be cautious with glass for cationic peptides. Siebels et al. found the highest number of adsorbed peptides on glass, driven by electrostatic interactions, and Zhou et al. found that treating a negatively charged membrane improved recovery only for peptides with net positive charge. Together these point to cationic peptides as the ones to watch on negatively charged surfaces such as glass; that reading is our inference from the two papers.
  3. Add a competitor or modifier where the assay allows. Albumin or Triton X-100 (Lindberg and Andersson, 1991), 0.05% plasma (Bongaerts et al., 2023), nonionic detergents (Kune et al., 2025) and acetonitrile (Zhou et al., 2015) each reduced loss in the system studied. Check that the additive does not interfere with your own assay; Vanderstichele et al. noted that detergent changed the measured Aβ concentrations.
  4. Dissolve properly first. Maes et al. (2014) describe proper solubilisation as one of the most important factors; see dissolving peptides for in-vitro assays.
  5. Limit transfers. The 73% loss in the serial-tube experiment came from six consecutive tubes; each additional surface adds loss.
  6. Prefer ratios or an internal standard where you can. Vanderstichele et al. found a ratio of two peptides more robust to pre-analytical factors than a single peptide.
  7. Control the LC system as well as the vial. Zhang et al. report that a C8 column, a higher flow rate, a slower gradient and a strong needle wash (0.2% formic acid in 90% acetonitrile) reduced carryover, together bringing adsorption of the susceptible peptides to about one-hundred-and-fiftieth of the starting level.

Interfaces beyond vials

Hamada et al. (2026) show that adsorption is not only a loss mechanism: for liraglutide in syringes, the interface with silicone oil together with an air-water interface promoted fibrillation. Headspace and agitation mattered in that study, which is one reason our guide to peptide stability in solution treats adsorption and aggregation as physical routes alongside chemical ones.

Related

References

  1. Maes K, Smolders I, Michotte Y, et al. (2014). Strategies to reduce aspecific adsorption of peptides and proteins in liquid chromatography-mass spectrometry based bioanalyses: an overview. J Chromatogr A 1358:1-13. PubMed 25022477 · DOI
  2. Lindberg BF, Andersson KE (1991). Adsorption of atrial natriuretic peptide to different materials: a factor influencing results of in vitro experiments?. Pharmacol Toxicol 68:276-81. PubMed 1830963 · DOI
  3. Siebels B, Moritz M, Hübler D, et al. (2024). Assay for Characterizing Adsorption-Properties of Surfaces (APS). Chemistry 30:e202403000. PubMed 39189660 · DOI
  4. Kune C, Tielens S, Baiwir D, et al. (2025). SIGNIFICANT IMPACT OF CONSUMABLE MATERIAL AND BUFFER COMPOSITION FOR LOW-CELL NUMBER PROTEOMIC SAMPLE PREPARATION. Anal Chem 97:3836-3845. PubMed 39933935 · DOI
  5. Zhang Y, Yang J, Ma Y, et al. (2022). [Nonspecific adsorption evaluation and general minimization strategy in peptide analysis based on ultra-performance liquid chromatography-mass spectrometry]. Se Pu 40:616-624. PubMed 35791600 · DOI
  6. Zhou Y, Wong JM, Mabrouk OS, et al. (2015). Reducing adsorption to improve recovery and in vivo detection of neuropeptides by microdialysis with LC-MS. Anal Chem 87:9802-9. PubMed 26351736 · DOI
  7. Bongaerts J, De Bundel D, Smolders I, et al. (2023). Improving the LC-MS/MS analysis of neuromedin U-8 and neuromedin S by minimizing their adsorption behavior and optimizing UHPLC and MS parameters. J Pharm Biomed Anal 228:115306. PubMed 36868028 · DOI
  8. Vanderstichele HM, Janelidze S, Demeyer L, et al. (2016). Optimized Standard Operating Procedures for the Analysis of Cerebrospinal Fluid Aβ42 and the Ratios of Aβ Isoforms Using Low Protein Binding Tubes. J Alzheimers Dis 53:1121-32. PubMed 27258423 · DOI
  9. Hamada N, Cutts A, Song J, et al. (2026). Molecular Mechanisms of Liraglutide Aggregation Induced by Dual Air-Water and Silicone-Oil-Water Interfacial Stress. Mol Pharm 23:1295-1309. PubMed 41572471 · DOI

Frequently asked questions

Do peptides really stick to plastic and glass?

Yes. Lindberg and Andersson (1991) measured adsorption of atrial natriuretic peptide to every material tested, including siliconized glass and polypropylene, and Siebels et al. (2024) found that many hydrophobic peptides adsorbed to conventional polypropylene vials and the most peptides adsorbed to glass.

Which peptides are most at risk?

The studies point to hydrophobic peptides on polypropylene, positively charged peptides on negatively charged surfaces such as glass, and larger neuropeptides. Zhang et al. (2022) found hydrophobicity index and length correlated with adsorption but explained only about 30% of the peptides, so treat any rule as a screen.

Does it matter at high concentration?

Much less. A surface that removes a fixed amount of peptide removes a large fraction of a dilute solution and a negligible fraction of a concentrated one; the table on this page shows the arithmetic.

What reduces adsorption?

Low-protein-binding polypropylene or vials made from polar polymers, avoiding glass for cationic peptides, an organic fraction or a competitor such as albumin, plasma or a nonionic detergent where the assay tolerates it, and fewer transfers between tubes.